Page #278 background image
Model
8901B
J.
Program
Code
HP FILTER
ALL
OFF
HO
50
Hz H1
300
Hz H2
Operation
Audio
Filters
Program
Code
LP FILTER
@+
ALL
OFF
LO
3
kHz L1
15
kHz L2
>20
kHz L3
FUNCTIONS
HP
FILTER,
LP
FILTER
DESCRIPTION
High-pass and low-pass filters can be inserted in the path of the demodulated
signal.
The filters are used for such
tasks
as rejecting interfering audio
signals
such
as
line hum and sub-carriers.
They
also
provide pre-defined bandwidths for residual noise measurements.
PROCEDURE
Insert the desired demodulated
signal
filter
by
selecting the appropriate keys. Only one high-pass and
one low-pass filter can be inserted
at
once.
To remove a filter, select the key again.
To
select a different filter, press the key corresponding
to
the desired filter.
HP-IB codes for the various filter keys (listed in the following table) only insert the
filters.
To remove
a high-pass or low-pass filter via
HP-IB,
use code
HO
or
LO
(respectively), or select another filter.
HP-IB PROGRAM CODES
Audio Filters
3-179

Table of Contents