HP4402.book : chapt5_2.fm 11 Tue Apr 22 15:20:23 1997
Specifications
Power Meter Supplemental Characteristics
HP EPM-441A User’s Guide 5-11
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
10 dB
10 dB
10 dB
60 dB
Maximum dBm
Minimum dBm
Power
Sensor
40 ms
3.4 s
210 ms
70 ms
Settling
Times
Typical
Dynam
ic
Range
10 dB
10 dB
10 dB
60 dB
Maximum dBm
Minimum dBm
Power
Sensor
70 ms
6.5 s
400 ms
120 ms
Settling
Times
Typical
Dynamic
Range
Normal Mode x2 Mode