Load Commands
Copyright © Itech Electronic Co., Ltd. 22
Returned value:
0|1
[SOURce:]PROTection:CLEar
This command is used to read the protection state when electronic load is
reset.
Command syntax:
[SOURce:]PROTection:CLEar
Parameter:
None
PROTection:AUTO:CLEar
This command is used to set the state when electronic load clear the protection
automatically.
Command syntax:
[SOURce:]PROTection:AUTO:CLEar[:STATe] <bool>
Parameter:
<0|1|OFF|ON>
Reset value:
0
Query Syntax:
[SOURce:]PROTection:AUTO:CLEar[:STATe]?
Returned value:
0|1
CURRent
This command is used to set the current value of electronic load.
Command syntax:
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>