M+/SAVE
to save.
4.
Put the corresponding weight on the weighing
and then press
TARE/
to complete the first point
Note:
After the first point calibration, the window can display the weight value. If no need
for the other point calibration, move to Step 6 to finish the calibration procedure.
5.
Add another weight 500g to the current weight. The window
will show the t
otal weights value.
Repeat this step to complete the calibrations of on3, on4 and on5.
6.
P
ress
ZERO/ESC
to
sa
v
e
.
A
It will return to normal weighing
Note: the window displays
return to the normal weighing.
7.
Operation
7-1 Weighing
Place item to be weighed on the scale. The Weight
window shown is 1.00kg, Gross Weight.
7-2 Tare & Preset Tare
Tare
When weighing a sample that must be held in a
into memory.
1)
Under the weighing mode, place the container on the scale.
Press the key
TARE
The
symbol appears and the window displays as shown on the
2
)
Place the item into the
container.
Then the window displays the net weight.
3
)
Clear the pan. The window displays the taring
That is the weight of
4) To
cancel tare action, clear the pan and press
or
ZERO/ESC.
(II)
Put the corresponding weight on the weighing
pan,
to complete the first point
calibration.
After the first point calibration, the window can display the weight value. If no need
for the other point calibration, move to Step 6 to finish the calibration procedure.
Add another weight 500g to the current weight. The window
otal weights value.
Press
TARE/
to complete.
Repeat this step to complete the calibrations of on3, on4 and on5.
ft
er
t
he
window
displa
ys
―
‖
,
It will return to normal weighing
mode.
automatically when
on5 is finished. It will be saved and
Place item to be weighed on the scale. The Weight
window shown is 1.00kg, Gross Weight.
When weighing a sample that must be held in a container, taring stores the container weight
Under the weighing mode, place the container on the scale.
to complete tare
action.
symbol appears and the window displays as shown on the
Then the window displays the net weight.
Clear the pan. The window displays the taring
weight
cancel tare action, clear the pan and press
TARE/
After the first point calibration, the window can display the weight value. If no need
for the other point calibration, move to Step 6 to finish the calibration procedure.
on5 is finished. It will be saved and
container, taring stores the container weight
right