58
SCPI Command Subsystems
Notes:
1. Brackets ([ ]) are used to denote optional character sets.
These optional characters do not have to be included in
the program message. Do not use brackets ([ ]) in the pro-
gram message.
2. Angle brackets (< >) are used to indicate parameter type.
Do not use angle brackets (< >) in the program message.
3. Upper case characters indicate the short-form version for
each command word.
Table 22. Signal oriented measurement commands
Command Description
:FETCh? Requests the latest reading (SCPI) or a
fresh reading (FRESh).
:CONFigure:<function> Places the 2001 in a “one-shot” mea-
surement mode for the specified func-
tion.
:READ? Performs an :ABORt, :INITiate, and a
FETCh?.
:MEASure[:<function>]? Performs an :ABORt, :CONFigure
:<function>, and a :READ?.
Table 23. Calculate command summary
Command Description
:CALCulate[1]
:FORMat <name>
:FORMat?
:KMATh
:MMFactor <NRf>
:MMFactor?
:MBFactor <NRf>
Subsystem to control CALC 1:
Select math format: MXB, PERCent,
NONE.
Query math format.
Path to configure math calculations:
Specify “m” for mX+b
(-9.999999e20 to
+9.999999e20).
Query “m” factor.
Specify “b” for mX+b
(-9.999999e30 to
+9.999999e30).