Page #63 background imageLoading...
Page #63 background image
2260B Series Programming Manual
60
[SOURce:]VOLTage:SLEW:RISing
Description
Sets or queries the rising voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
Query Syntax
[SOURce:]VOLTage:SLEW:RISing {<NRf>|MIN|MAX}
[SOURce:]VOLTage:SLEW:RISing? [MIN|MAX]
Parameter/Return <NRf>
0.01V/s--200% (2260B-30-XX)
0.1V/s--200% (2260B-80-XX)
MIN
Minimum rising voltage slew rate.
MAX
Maximum rising voltage slew rate.
Example SOUR:VOLT:SLEW:RIS MAX
Sets the rising voltage slew rate to its maximum.
[SOURce:]VOLTage:SLEW:FALLing
Description
Sets or queries the falling voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax [SOURce:]VOLTage:SLEW:FALLing {<NRf>|MIN|MAX}
Query Syntax [SOURce:]VOLTage:SLEW:FALLing? [MIN|MAX]
Parameter/Return
<NRf>
0.01V/s--200% (2260B-30-XX)
0.1V/s--200% (2260B-XX)
MIN
Minimum voltage falling slew rate.
MAX
Maximum voltage falling slew rate.
Example SOUR:VOLT:SLEW:FALL MIN
Sets the falling voltage slew rate to its minimum.
Set
Query
Set
Query

Table of Contents

Other manuals for Keithley 2260B-30-36

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-30-36 and is the answer not in the manual?

Keithley 2260B-30-36 Specifications

General IconGeneral
BrandKeithley
Model2260B-30-36
CategoryPower Supply
LanguageEnglish

Related product manuals