REMOTE CONTROL
65
Return parameter
<Boolean>
Returns the buzzer status.
SYSTem:CONFigure:BLEeder[:STATe]
Description
Sets or queries the status of the bleeder resistor.
Syntax
Query Syntax
SYSTem:CONFigure:BLEeder[:STATe] {OFF|ON|0|1}
SYSTem:CONFigure:BLEeder[:STATe]?
Parameter 0
<NR1> Turns the bleeder resistor off.
OFF
Turns the bleeder resistor off.
1
<NR1> Turns the bleeder resistor on.
ON
Turns the bleeder resistor on.
Return parameter
<Boolean>
Returns bleeder resistor status.
SYSTem:CONFigure:BTRip[:IMMediate]
Description
Trips the power switch trip (circuit breaker) to
turn the unit off (shut down the power).
Syntax SYSTem:CONFigure:BTRip[:IMMediate]
SYSTem:CONFigure:BTRip:PROTection
Description
Enables/Disables the power switch trip (circuit
breaker) when the OVP or OCP protection settings
are tripped. This setting only applies after power
has been cycled.
Syntax SYSTem:CONFigure:BTRip:PROTection
{DISable|ENABle|0|1}
Query Syntax SYSTem:CONFigure:BTRip:PROTection?
Parameter
0
<NR1> Enables the power switch trip for
OVP or OCP.
DISable
Disables the power switch trip for OVP or
OCP.
1
<NR1> Disables the power switch trip for
OVP or OCP.
ENABle
Enables the power switch trip for OVP or
OCP.
Set
Query
Set
Set
Query

Table of Contents

Other manuals for Keithley 2260B-30-36

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-30-36 and is the answer not in the manual?

Keithley 2260B-30-36 Specifications

General IconGeneral
BrandKeithley
Model2260B-30-36
CategoryPower Supply
LanguageEnglish

Related product manuals