2260B Series Programming Manual
62
Example:
Immediate Mode
TRIG:TRAN:SOUR IMM
CURR:TRIG MAX
VOLT:TRIG 5
INIT:NAME TRAN
The current changes to the maximum, and the voltage
changes to 5V.
Example:
Bus Mode
TRIG:TRAN:SOUR BUS
CURR:TRIG MAX
VOLT:TRIG 5
INIT:NAME TRAN
TRIG:TRAN
The current changes to the maximum, and the voltage
changes to 5V.
TRIGger:OUTPut[:IMMediate]
Description
Generates a software trigger for the output trigger
system.
Syntax TRIGger:OUTPut[:IMMediate]
TRIGger:OUTPut:SOURce
Description
Sets or queries the trigger source for the output
system.
Syntax TRIGger:OUTPut:SOURce [BUS | IMMediate]
Query Syntax TRIGger:OUTPut:SOURce?
Parameter/Return
BUS
Internal software trigger. Waits for the
*TRG (or IEEE 488.1 “get” group
execute trigger) command to start the
trigger.
IMMediate
Starts the trigger immediately.
(default)
Set
Set
Query