Quick Results Guide 7
4-wire connections
Connections for 4-wire local sensing are shown in Figure 4. Use 4-wire remote sensing for
the following source-measure conditions:
•Test circuit impedance is <1k
.
Optimum ohms, V-source, and/or V-measure accuracy is required.
With 4-wire sensing selected, the 4W annunciator is on. The following menu sequence
selects 4-wire remote sensing:
Press CONFIG > select SOURCE V (or MEAS
) > select SENSE MODE > select 4-WIRE
NOTE
Specified accuracies for both source and measure are achieved only using 4-wire
remote sensing.
HI
LO
OUTPUT
INPUT/
OUTPUT
4-WIRE
SENSE
ON/OFF
TERMINALS
FRONT/
REAR
DUT
SourceMeter Front Panel Sense Selection: 4-wire
FRONT/
REAR
Figure 4
4-wire connections