EasyManuals Logo
Home>Keithley>Measuring Instruments>SourceMeter 2400

Keithley SourceMeter 2400 Quick Results Guide

Keithley SourceMeter 2400
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Quick Results Guide 9
NOTE
Ohms guard cannot be used for high current (>100mA) ranges. Ohms guard cannot
be selected if already on a high current range. Conversely, if ohms guard is already
selected, a high current range cannot be selected.
The following menu sequence selects ohms guard:
Press CONFIG > press MEAS
(or SOURCE V or SOURCE I) > select GUARD > select
OHMS
Basic connections for ohms guard is shown in Figure 6A. With V,
Guard and In/Out HI at
virtually the same potential, the voltage drop across R will be 0V, and therefore, the current
through R will be 0A. Virtually all the current from In/Out HI will flow through the DUT
resulting in an accurate measurement.
If the guard resistance path (R
G
) is <1k
, IR drop in the V,
Guard test lead could be high
enough such that a significant voltage drop appears across R. The resultant leakage current
through R will corrupt the measurement of the DUT.
Guard Sense is used to cancel the effect of IR drop in the V,
Guard test lead as shown in
Figure 6B. Guard Sense regulates the guard voltage to ensure that the same potential appears
on either side of R.
For DUT <1k
, 4-wire sensing should be used as shown in Figure 6C.
NOTE
Guard current (I
G
) must never exceed 50mA. If it does, guard voltage will become
less than the output voltage and corrupt the measurement.
DUT
V, Guard
Sense HI
In/Out HI
In/Out LO
Sense LO
Guard Sense
Sense Mode: 4-wire
C. 6-wire connections
DUT
V, Guard
In/Out HI
In/Out LO
Sense Mode: 2-wire
A. Basic connections
DUT
V, Guard
In/Out HI
In/Out LO
Guard Sense
Sense Mode: 2-wire
B. Connections using guard sense
<1k
RG
IG
1k
RG
IG
R
0V
R
0V
Figure 6
Guarded ohms measurements

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals