26 Quick Results Guide
Logarithmic staircase sweep
This sweep is similar to the linear staircase sweep. The steps, however, are done on a loga-
rithmic scale. The symmetrical log points for the steps are determined by the specified number
of sweep points. Figure 8 shows a 5-point log sweep from 1 to 10V.
With the desired source (V or I) selected, use the following menu sequence to set the start
and stop levels, and the number of sweep points (2 to 2500):
Press CONFIG > press SWEEP > select TYPE > select LOG
After selecting LOG, you will be prompted to enter the START and STOP levels, and the
NO OF POINTS.
Delay
Delay
Delay
Delay
Start
Bias
Stop
(10)
Measure
#1
Measure
#2
Measure
#3
Measure
#5
X = Measurement Point
X
X
X
X
Delay
X
Measure
#4
Log Points = 5
1
10
5.6234
3.1623
1.7783
Volts
Log
Scale
Figure 8
Logarithmic staircase
sweep (example)

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals