22 Quick Results Guide
Remote command programming
The SCPI commands for speed, digits, filter, and rel are listed in Table 12. The commands for
ranging are listed in Tables 6 and 7.
Table 12
SCPI commands; speed, digits, filter, and rel
Command Description
Speed commands
1
:
[:SENSe]:CURRent:NPLCycles <n> Set measurement speed; <n> = 0.01 to 10.
[:SENSe]:VOLTage:NPLCycles <n> Set measurement speed; <n> = 0.01 to 10.
[:SENSe]:RESistance:NPLCycles <n> Set measurement speed; <n> = 0.01 to 10.
Digits command:
:DISPlay:DIGits <n> Set display resolution; <n> = 4, 5, 6, or 7.
Filter commands:
[:SENSe]:AVERage:TCONtrol <name> Select filter type; <name> = REPeat or MOVing.
[:SENSe]:AVERage:COUNt <n> Set filter count; <n> = 1 to 100.
[:SENSe]:AVERage <b> Enable/disable filter; <b> = ON or OFF.
Rel command
2
:
:CALCulate2:NULL:STATe <b> Enable/disable rel; <b> = ON or OFF.
1. The speed setting is global for all functions. Therefore, you can use any of the three commands to set speed.
2. The commands to acquire a rel value are not listed in this table (see Section 8, Relative in the 2400 Series
SourceMeter User’s Manual).

Table of Contents

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals