440-BA-e-1744
Using this function data are selected which are to be sent via the
RS232C (not valid for data transfer mode BAPr ).
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE key several times until „LAPr“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired output parameter by pressing the MODE key
Edition of the total weight
Edition of the net weight
Edition of the tare weight
Edition of the stored weight
Edition of the unit weight
Edition of the reference quantity
Edition of a page feeding at start printer output
Edition of a page feeding at end printer output
After actuating the SET button, the current state is displayed ( on
/ off ).
Use MODE and PRINT key to change the status
„on off“.
Use the SET key to confirm selection. The balance returns to
weighing mode.
By that way the user can configurate his own data block, which then
is sent to a printer or to a PC.
Baud rate
The baud rate defines the transfer speed vie the interface, 1 Baud =
1 Bit/second.
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE key several times until „bAUd“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Use MODE key select the desired settings
9600 4800 2400 1200 19200
Use the SET key to confirm selection. The balance returns to
weighing mode.