4.2.1 F1 - CALIBRATION
To enter the calibration parameter, press the [Tare] key when “F1” is displayed.
The display will instruct you to remove any weight from the scale by displaying “UnloAd
Press the [Tare] key.
The display will tell you to add the calibration weight to the scale: “LoAd 1” “ 05
kiLo5
Add the weight shown or press the key to clear the value and enter another value
of weight. Wait for the stability indicator and then press the [Tare] key.
The display will show “SPAn” “PASS” if the calibration is OK or it will show “SPAn” “FAIL
if calibration could not be completed correctly.
The display will show “JP On”. Remove the calibration weight.
Remove the jumper and the scale will run a self-test and return to normal operation.
4.2.2 F2 - RESOLUTION
This function is not active on the CPB-M series.
4.2.3 F3 - CAPACITY
To enter this parameter, press the [Tare] key when “F3” is shown.
The display will show the current capacity in 3, 6, 15 or 30 kilograms.
Press the key to change the value.
Press [Tare] to accept the displayed value.
NOTE: The scale must be originally built for the capacity selected. The 3kg unit
uses the 5kg load cells, 6kg unit uses 10kg load cells, the 15kg unit uses 20kg load
cells and the 30kg unit uses 35kg load cells.
CPB-SH-e-0721 12