4,2.7 F7 - A/D COUNT
To enter this parameter, press the [Tare] key when “F7” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This
can be an aid to service.
Press the [Tare] key to return to the PARAMETER menu.
Press the key to return to normal weighing.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)
4.2.8 F8 - ZERO
The scale can be selected to use a normal method to determine zero, “Mode 1” or an
alternative method “Mode 2”. For all CPB-M scales set this option to “Mode 1”.
To enter this parameter, press the [Tare] key when “F8” is shown.
The display will show the selected zero mode.
Press the
key to change the value.
Press [Tare] to accept “
Mode 1”.
4.2.9 F9 - PIN
To change the user Pin number to gain access to the Service Parameters.
To enter this parameter, press the [Tare] key when “F9” is shown.
The display will show “Pin 1
Enter the new Pin number. Press the [Tare] key to accept.
CPB-SH-e-0721 14