EasyManuals Logo
Home>KERN>Scales>EG 620-3NM

KERN EG 620-3NM User Manual

KERN EG 620-3NM
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
EW/EG-N/EWB-e-1627
26
5.10 Verification switch and official seal
Prior to verification the verification switch must be moved from the displayed position
(see direction of arrow) to verification position. Once in this position, a parenthesis
will be displayed around the last display point.
Following calibration the balance is sealed at the marked position.
Balance calibration is not valid without the “official seals“.
Position of the “official seals“:
Move verification switch
Balances that are subject to compulsory verification must be taken out of op-
eration if:
The weighing result of the balance is outside the error limit. Therefore, apply a
known test weight (approx. 1/3 of the nominal load) to the balance at regular in-
tervals and compare with the display value.
The reverification deadline has been exceeded.
Official seal
Official seal

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EG 620-3NM and is the answer not in the manual?

KERN EG 620-3NM Specifications

General IconGeneral
BrandKERN
ModelEG 620-3NM
CategoryScales
LanguageEnglish

Related product manuals