8.5 Data input commands
8.5.1 Command input format
is made up of 4 characters, CR=0DH, LF=0AH
8.5.3 External control commands
C1 C2 Code Meaning
O 0 4FH 30H No data output
O 1 4FH 31H Continuous data output
O 2 4FH 32H Continuous output of stable weight value data
O 3 4FH 33H
Output of stable and unstable weight values after pressing the
PRINT button
O 4 4FH 34H
Data output upon stable weight value, having previously unloaded
the weighing scale
O 5 4FH 35H
Data output upon stable weight value. No data output at unstable
weight value. Data output again, after stabilization has been
reached
O 6 4FH 36H
Data output upon stable weight value. Continuous data outputting
at unstable weight value.
O 7 4FH 37H Output of stable weight values after pressing the PRINT button
O 8 4FH 38H One-time immediate data output*
O 9 4FH 39H One-time data output after stabilization has been reached*
O A 4FH 41H
One-time, immediate data output after lapse of a defined period of
time*
O B 4FH 42H
One-time, immediate data output after lapse of a defined period of
time and upon stable weight value*
* While using these commands for remote control purposes, do not press the PRINT button (data
transmission error). In case data transmission error occurs, briefly cut off the weighing scale from its
power supply source.