7.6 Temperature calibration / adjustment
The temperature can be checked and re-adjusted if necessary using the optional
temperature calibration set (MLB-A03).
Temperature calibration
-
The balance is in normal weighing mode.
-
Insert the temperature sensor through the balance cover opening and push in
close to the weighing plate (see pictures).
Guide bush for temperature sensor
-
-
-
Press the key and then press the and keys simultaneously.
A “Code“ is now queried on the display screen.
-
Use the arrow keys to enter the correct code (“002003“).
Numerical entry from 0-9
For changing the entry point on the display screen
-
Press the key. “St_tE “ will appear on the display screen - the temperature
sensor setting programme.
-
Press the key. The temperature inside the weighing region will now appear.
If this is not 20°C, please use the arrow keys to alter the temperature.
-
Press the key. If a temperature of 20°C is entered, a number of approxi-
mately 4000 will appear on the display screen.
-
The key is now used to start temperature calibration. The balance will now
heat up to 50% of its total power.