EasyManuals Logo
Home>KERN>Scales>PFB Series

KERN PFB Series User Manual

KERN PFB Series
Go to English
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
PFB-BA-e-1630 15
„Pass“ appears, the adjustment process has been
finished successfully.
Remove adjusting weight during selftest
Wait until the balance is again in weighing mode.
5.8 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
In balances with a resolution of > 15 000 dividing steps carrying out a
linearization is recommended.
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter “testing instruments control”.
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
After successful linearization you will have to carry out calibration;
see chapter “testing instruments control”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PFB Series and is the answer not in the manual?

KERN PFB Series Specifications

General IconGeneral
BrandKERN
ModelPFB Series
CategoryScales
LanguageEnglish

Related product manuals