EasyManuals Logo
Home>KERN>Scales>TCKE 300-3-A

KERN TCKE 300-3-A Operating Instruction

KERN TCKE 300-3-A
Go to English
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
38 CKE-BA-e-2132
13.2.3 Automatic data transfer <auto>
Data transfer takes place automatically without pressing the PRINT button provided the ap-
propriate transfer conditions are met depending on the menu setting.
Function enabling and transfer condition setting:
In the setup menu, display the <Print PrMode> menu item and confirm,
pressing .
To transfer data automatically, using navigation buttons , choose the <auto> menu
item and confirm by pressing .
Use the navigation buttons to choose the setting <on> and confirm by pressing .
The <zRange> symbol will be displayed.
Confirm by pressing and use the navigation settings to set the required transfer
condition.
Confirm pressing .
To leave the menu, press the navigation button several times.
Placing the weighed material:
Whenever required, place an empty container on the scale and tare it.
Place the weighed material and wait until the stabilization indicator is displayed ( ).
The weighing value will be transferred automatically.
13.2.4 Continuous data transfer <cont>
Function enabling and transfer cycle setting:
In the setup menu, display the <Print PrMode> menu item and confirm,
pressing .
To transfer data continuously, using navigation buttons , choose the <cont> menu
item and confirm by pressing .
Use the navigation buttons to choose the setting <on> and confirm by pressing .
The <speed> symbol will be displayed.
Confirm by pressing and using the navigation buttons, , set the required cycle (for
introducing the numerical value, see chapter 3.2.2).
To leave the menu, press the navigation button several times.
Place the weighed material.
Whenever required, place an empty container on the scale and tare it.
Place the weighed material.
The weighing values will be transferred in line with the predefined cycle.

Table of Contents

Other manuals for KERN TCKE 300-3-A