7. Error Messages
No. Error message Counter measure / causes
E99 System Error, Contact
Services !
Hardware faults or file errors or adjustment errors or setting errors
rendering instrument not working properly.
E100 Low Battery ! Change to new or freshly charged batteries.
E102 Too Bright ! Darken staff or reduce lighting on staff or shade the objective telescope.
E103 Too Dark ! Light up staff uniformly.
E104 No Staff ! Check targeting.
E106 Out Of Level ! Level the instrument.
E108 Data File Error ! Data file error.
E110 Target Too Close ! Move staff or instrument further apart.
E111 Target Too far ! Move staff or instrument closer together.
E112 Too Cold ! Stop working, external temperature is outside the instrument operating
temperature.
E113 Too Hot ! Stop working, external temperature is outside the instrument operating
temperature.
E114 Invalid Measurement ! Make another measurement. If further measurement proved to be futile,
check staff position and Inverse Staff setting, check the lighting condition
at the staff and stray light, check focusing and targeting, check if sufficient
length of barcode in the field of view.