19.3.5 Electric Control
Plug connections:
► Replace O-rings on connectors.
► Check plugs connections for dirt and defects.
Clean or replace if necessary.
► Check that the plug connectors are properly seated.
Motors
► Check motor movement (adjust bed positions).
Check for incorrect and interrupted movements.
Have defective motors replaced if necessary.
► Check cables for signs of wear and entanglement.
Install a new cable or have it replaced if necessary.
Accumulator
► Check that the accumulator is working properly (disconnect the bed from the mains).
Have the accumulator replaced if necessary.
Fuses
► Have fuses changed exclusively by qualied and trained service technicians authorised by the manufacturer.
► Use the following fuse types exclusively:
■ T1.6 A (for 230 V input)
■ T3.15 A (for 100 - 127 V input)
19.3.6 Castors
► Clean the castors completely.
► Grease the castors if necessary (Caro EP 2 by DEA or an equivalent grease).
► Check that the castors work properly.
■ Forward Movement
■ Unrestricted Movement
■ Braked
► Have the brakes adjusted if necessary.
► Have any defective castors replaced.
19.3.7 Accessories
► Check that all accessories (for example, lifting pole, siderails, infusion stand, etc.) are working properly.
► Replace if necessary.
19.4 Safety Technical Checks
WARNING!
Risk of injury due to incorrect safety checks!
► Ensure that safety checks are performed exclusively by customer service or authorised personnel (certied by
the manufacturer).
► Ensure that the safety checks are recorded in the service and maintenance log.
WARNING!
Risk of injury due to defective bed!
► Have a defective bed repaired immediately.
► If the defect cannot be repaired, do not continue to use the bed.
Technical Safety Check of the medical bed must be performed at least once every 12 months.
The procedure for performing the safety check is stipulated in EN 62353:2014.
NOTE LINET ® provides service documentation for qualied personnel.