! CAUTION
Provision of information on the avoidance of light hazard from the optical device is required in
ISO 15004-2:2007
“Ophthalmic instruments-Fundamental requirements and test methods”
1. The manufacturer shall, on request, provide the user with a graph showing the relative spectral
output of the instrument between 305 nm and 1 100 nm when the instrument is operating at
maximum light intensity and maximum aperture. The spectral output shall be shown for the beam
after it exits the instrument.
LRK-7000
< Spectrum output of all light source during measurement (maximum light intensity) >
2. “CAUTION – The light emitted from this instrument is potentially hazardous. The longer the
duration of exposure, the greater the risk of ocular damage. Exposure to light from this instrument
when operated at maximum intensity will exceed the safety guideline after 1.72 minutes.”