3-33
ECG waves are displayed correctly without noise and the HR
value is within 80±1 bpm.
Yes N o
ECG Lead Off alarm behaves correctly. Yes No
Paced signals are detected and pace pulse marks are displayed
when [Paced] is set to [Ye s ]
Yes N o
The difference between the amplitude of the ECG calibration
square wave and that of the wave scale is not greater than 5%.
Yes N o
Resp test
The Resp wave is not distorted and the Resp value is within
40±2 rpm.
Yes N o
SpO
2
test
Measure SpO
2
on a healthy person’s finger and a Pleth wave and
PR value are displayed. The displayed SpO
2
value is within 95%
and 100%
Yes N o
NIBP test and calibration
The difference is within ±3 mm when 0, 50 or 200 mmHg is set
for NIBP accuracy test.
Yes N o
There is no leakage with NIBP, or the manual leakage test result
does not exceed 6mmHg/min.
Yes N o
Temp test
The value displayed for each Temp channel of the monitor is
within 37±0.1ºC.
Yes N o
IBP test
The static pressure value displayed for each IBP channel is
within 200±2 mmHg.
Yes N o
The ART and LV waves for each IBP channel are displayed
correctly.
Yes N o
C.O. test
The TB value displayed on the monitor is within 37±0.2ºC. Yes No
The displayed C.O. value is within 5±0.25L/min. Yes No
Mainstream CO
2
test
The mainstream CO2 is zeroed successfully and the waveform
baseline recovers to zero.
Yes N o
CO2 Apnea alarm behaves correctly. Yes No
The displayed CO2 value is within 6.0±0.3%.
Yes N o
Sidestream CO
2