2. Unscrew the four M3×6 screws underneath the screen cover and then remove the LCD.
5.3.9 Removing the Alarm LED Board
1. After removing the LCD, disconnect the cable that connects the alarm LED board and
the button board, and then, unscrew the two PT2×6 screws to remove the alarm LED
board.