EasyManuals Logo
Home>Mindray>Medical Equipment>P8-2Ts

Mindray P8-2Ts User Manual

Mindray P8-2Ts
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Examination
Operator's Manual 29
CAUTION
To avoid damaging the transducer, do not deflect the distal tip of the transducer by
direct application of force. Use the deflection wheels for this task.
4.3.3 Emergency Retraction
Ifthetransducertipshouldgetjammedinadeectedpositioninsidethepatient,andall
attemptstoreleasethedeectedtipshouldfail,followthefollowingproceduretoassurea
saferetractionofthetransducer.
1. Disconnectthetransducerfromtheultrasoundsystem.
2. Atanaccessiblelocationbetweenthetransducerhandleandthepatient,cuttheentire
endoscopeshaft,includingallinternalwiring,usingheavydutycuttingpliersoranother
suitable tool.
Thedeectionmechanismisnowreleasedandthetransducermaybesafelyretracted.
4.4 After Examinations
Aftertheexaminationiscompleted,turnOFFtheultrasonicdiagnosticsystem,removethe
bite-guardanddisconnectthetransducer,andthencleananddisinfectthetransducer.
Afterdisinfectingthetransducer,conrmthatthetransducerisingoodconditionandstoreit
inasuitableplace.
Fordetailsaboutcleananddisinfect,see"5 CleaningandDisinfection".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray P8-2Ts and is the answer not in the manual?

Mindray P8-2Ts Specifications

General IconGeneral
BrandMindray
ModelP8-2Ts
CategoryMedical Equipment
LanguageEnglish

Related product manuals