CleaningandDisinfection
34 Operator's Manual
5.3.2 Disinfection Procedure
1. Selectanappropriatehigh-leveldisinfectanttodisinfecttheprobe.Followthe
manufacturer'sinstructionstousethehigh-leveldisinfectant.Prepareadisinfectantby
usingsteriledistilledorsoftenedwaterwhennecessary.
2. Selectanappropriatemethod:
Soaking:Soaktheprobeinthedisinfectantsolutionandshaketheprobeproperly
toremovebubblesonthesurfaceoftheprobe.Fortheprobesoakingduration,see
thedisinfectantmanufacturer'sinstructions.
Onlythesectionfromdistalendtothe100cmmarker (70cmforpediatricTEE
probe)ontheflexibleshaftcanbeimmersedinthedisinfectantsolution.The
handle,cableandconnectorcanonlybedisinfectedwithawrungoutsoftcloth
moistenedwithdisinfectants.
Wiping:Useacommerciallyavailabledisinfectingwipe,oradisposablesterilelint-
freesoftclothsoakedwithadisinfectingspraytowipeanddisinfectallsurfacesof
theprobethoroughlyforthetimespeciedinthemanufacturer'sinstructions.
3. Rinsetheprobeinsertionpartthoroughlywithplentyofcleanowingwater(about7
.
5L)
atroomtemperatureforabout1minutetoremovetheresidualdirtanddisinfectants.
Orfollowtherinsingmethodspeciedbythemanufacturer.Usemoistenedsoftcloth
towipetheresidualdisinfectantsonthehandle,cable,andconnector.
4. Drytheprobewithadisposablelint-freesoftclothortissue.Donotdrytheprobeby
heating.
5. Checkwhethertheprobehasdefectssuchaspeeling,rifts,bumps,cracks,orliquid
spill.Ifsuchdefectsexist,theprobehasreachedtheendofitsservicelife.Inthis
case,stopusingitandcontacttheMindrayservicedepartment.
6. Storetheprobeinacoolanddryenvironment.
CAUTION
The lens may be discolored; the label on the transducer may fade. These are not
abnormalities.
Repeated disinfection will eventually damage the transducer, please check the
transducers performance periodically.
Disinfecting incorrectly or with chemicals not recommended by Mindray will void
the warranty.
WARNING
To avoid injury to the patient, if residue from the disinfectant is not removed it can
cause irritation and/or burning of the mouth and esophageal tissue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray P8-2Ts and is the answer not in the manual?

Mindray P8-2Ts Specifications

General IconGeneral
BrandMindray
ModelP8-2Ts
CategoryMedical Equipment
LanguageEnglish

Related product manuals