CleaningandDisinfection
Operator's Manual 31
5
Cleaning and Disinfection
ThissectiondescribesthemethodsandprecautionsforcleaninganddisinfectionofTEE
probes.Aftercompletingeachexamination,clean,disinfecttheTEEprobesasrequired.If
necessary,repeatthecleaning,disinfectionprocessbeforenextuse.
WARNING
Keep the control handle and system connector out of any cleaning or disinfection
solutions. The control handle and cable may be cleaned with a damp cloth, but
only the distal end of the transducer up to the 100 cm marker on the shaft (70 cm
for P8-3Ts) may be placed into a disinfection solution.
Do not use other disinfection methods like Iodine, Steam, Heat or Ethylene Oxide.
To avoid injury to the patient, you must follow the manufacturer’s
recommendation for rinsing.
CAUTION
To avoid damaging the transducer, the transducer should not be exposed to the
disinfectant longer than specified to achieve the desired effect, but never longer
than one hour.
To avoid damaging the transducer, do not steam autoclave or subject the
transducer to Ethylene Oxide (ETO).
To avoid damaging the transducer, do not immerse the transducer in a solution
containing ethanol.
After disinfection, rinse the transducer thoroughly with clean water to remove all
chemical residues. Chemical residues on the transducer may be harmful to the
human body.
The efficacy of disinfectants solutions is not guaranteed by MINDRAY. Contact
the manufacturers for information on the activity of the products.
NOTE:
Aftertheexamination,wipeofftheultrasoundgelthoroughly,otherwise,theultrasound
gelmaysolidifyanddegradetheimagequalityofthetransducer.
Donotpermitthetransducertobecomeoverheated(morethan55°C)duringcleaning
anddisinfections.Hightemperaturemaycausethetransducertobecomedeformedor
damaged.
Cleantheprobethoroughlyinaccordancewiththecleaningprocedurebefore
disinfection.