1-2 Passport 8/12 Service Manual
1.1.1 Warnings
WARNING
All installation operations, expansions, changes, modifications and repairs of this product are conducted by
Mindray authorized personnel.
There is high voltage inside the equipment. Never disassemble the equipment before it is disconnected
from the AC power source or the battery.
When you disassemble/reassemble a parameter module, a patient leakage current test must be performed
before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with protective earth contacts only.
If the installation does not provide for a protective earth conductor, disconnect it from the power line and
operate it on battery power, if possible.
Disposal of the packaging material should observe the applicable waste control regulations. and keeping it
out of children’s reach.
1.1.2 Cautions
CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the equipment when
preparing to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible
source of interference as they may emit higher levels of electromagnetic radiation.
Before connecting the receiver to the power line, check that the voltage and frequency ratings of the power
line are the same as those indicated on the unit’s label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force
during servicing.
1.1.3 Notes
NOTE
Refer to Operator’s Manual for detailed operation and other information.