Or
a. Select Alarms tab from Navigation Area
b. Select Limits
c. Select alarm limit group's tab (i.e. ECG/Resp, Oximetry, etc.)
d. Select parameter
e. Select High or Low limit value
f. Adjust value using on screen keypad or select Off to disable the
parameter’s alarm limit(s)
g. Select Enter
h. Select Accept to save settings and return to the Alarm Limits tab
i. Select Accept to return to the main screen
Panorama® Central Station interface: Changing alarm limits on the Panorama
Central Station or WorkStation will automatically change alarm limits on the
V12 or V21 bedside monitor. Changing alarm limits at the V12 or V21 bedside
monitor will automatically change alarm limits on the Panorama Central
Station and WorkStation.
Adjust Alarm Responses
To adjust a patient’s alarm responses on the V12 or V21 bedside monitor:
a. Select Alarms tab from Navigation Area
b. Select Responses
c. Select applicable alarm limit group's tab (i.e. ECG/Resp, Oximetry, etc.)
d. Select parameter
e. Select
• Alarm Level – to adjust the audio and visual alarm indicators.
Level options include High (Red), Medium (Yellow), and Low (Blue)*
• Print on Alarm – to activate an automatic printing when the alarm
occurs
• Save to Event – to save a 20 second event in the Events tab
(History menu)
(A in the box indicates selected)
f. Select Accept to save settings and return to the main screen