EasyManuals Logo
Home>National Instruments>Control Unit>5761R

National Instruments 5761R User Manual

National Instruments 5761R
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
NI 5761R User Guide and Specifications 14 ni.com
Table 5 lists the AC-coupled spectral performance measurements. All values are measured with a
250 MHz external Sample clock.
Channel to channel isolation
1 MHz ............................................................>90 dB
100.1 MHz .....................................................90 dB
501 MHz ........................................................80 dB
AC-Coupled Measurements
Figure 8. AC-Coupled Bandwidth (Passband)
Table 5. AC-Coupled Spectral Performance
Measurement 20.17 MHz 70.17 MHz
*
123.17 MHz
*
SNR 72.5 dB 71.4 dB 70.5 dB
SINAD 72.3 dB 71.2 dB 70.3 dB
SFDR 88 dB 84 dB 80 dB
*
These measurements were extrapolated from a –4 dBFS plot.
1M80k
–15
–14
–13
–12
–11
–10
–9
–8
–7
–6
–5
–4
–3
–2
–1
0
1
2
3
4
5
Gain (dB)
Frequency (Hz)
100k 10M 100M 700M