Physical Characteristics
Dimensions 16.0 cm × 10.0 cm (6.3 in. × 3.9 in.), 3U
Connector
GPIB IEEE 488 24-pin connector
PXI 32-bit, Universal Signaling
Environment
Operating ambient temperature 0 °C to 55 °C
(Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2.)
Operating relative humidity 10% to 90%, noncondensing
(Tested in accordance with IEC 60068-2-56.)
Storage ambient temperature -20 °C to 70 °C
(Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2.)
Storage relative humidity 5% to 95%, noncondensing
(Tested in accordance with IEC 60068-2-56.)
Shock and Vibration
Shock
Operating 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27;
test profile developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating 5 Hz to 500 Hz, 0.3 g
rms
Nonoperating 5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC 60068-2-64.
Nonoperating test profile exceeds the
requirements of MIL-PRF-28800F, Class 3.)

Other manuals for National Instruments GPIB-USB-HS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments GPIB-USB-HS and is the answer not in the manual?

National Instruments GPIB-USB-HS Specifications

General IconGeneral
BrandNational Instruments
ModelGPIB-USB-HS
CategoryComputer Hardware
LanguageEnglish

Related product manuals