Chapter 2 Theory of Operation
PCI E Series RLPM 2-16
©
National Instruments Corporation
Figure 2-12. Multirate Scanning of Two Channels with 3:1:1 Sampling Rate
Here, channel 0 is sampled three times, whereas channels 1 and 2 are
sampled once every three scans.
Figure 2-13. Multirate Scanning of Three Channels with 4:2:1 Sampling Rate
Multirate Scanning Using Ghost
If the ghost option in the configuration memory is set, that conversion
occurs but the data is not stored in the analog input FIFO. In other words,
a conversion is performed and the data is thrown away. By using this
option, multirate scanning with ratios such as x:y are possible, within the
limits imposed by the size of the configuration memory.
Figures 2-14 and 2-16 illustrate the advantages of using the ghost feature.
Figure 2-15 shows Example 3 timing, and Figure 2-16 shows the same
example using ghost.
Example 3: channel 1: channel 0 = 2:3 (without ghost).
START
STOP
Sampling Rates
Channel 0: Channel 1: Channel 2 = 3:1:1
CONVERT
0 0 01 2
START
STOP
CONVERT
0
000121
Sampling Rates
Channel 0: Channel 1: Channel 2 = 4:2:1

Table of Contents

Other manuals for National Instruments PCI-6035E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PCI-6035E and is the answer not in the manual?

National Instruments PCI-6035E Specifications

General IconGeneral
BrandNational Instruments
ModelPCI-6035E
CategoryComputer Hardware
LanguageEnglish

Related product manuals