26 | ni.com | NI PXI-4130 Calibration Procedure
5. Record the NI 4130 output voltage measurements for both external voltages.
6. Calculate the Voltage Change Limit using the formula in Table 13.
7. Subtract the two voltage measurements, V
1
- V
2
, to calculate the Voltage Change. The test
passes if the Voltage Change falls within the calculated Voltage Change Limit.
Table 13. NI PXI-4130 Voltage Line Regulation Voltage Change Limit Formulas
Channel Voltage Change Limit
1 ± ((V
1
× 1.00 × 10
-4
) + 1.00 × 10
-3
) × (V
ext1
- V
ext2
)