Note: The “X” key with the “pause” icon allows you to pause screening
from the Screening in Progress or Infant Data Window (SpeedScreen only).
Pause and Cancel Screening
During the screening, you may pause screening at any time by clicking the X key.
The display will indicate “Screening Paused” and will wait indefinitely. To resume
screening, click OK.
To cancel screening, click X again. The ALGO 3i screener will ask for confirmation
that you want to exit the screening. If a result for at least one ear has been obtained,
the ALGO 3i screener will save and confirm a screening record. If neither ear has
obtained a result, no screening record will be saved.
There are a few conditions under which the ALGO 3i screener will automatically
terminate the screening process:
• Cable(s) disconnected
• Battery charger connected
• Screening time limit exceeded
• Battery too low to complete screening
Step 8. Observe screening complete: At the completion of the screen, the ALGO 3i screener
will save a screening record to memory, and will display the infant data, including a
summary of the screening results.
Step 9. Remove/discard supplies: Remove from the baby and discard the Natus screening
supplies.
Step 10. Return to MainMenu/Disposition screening record: Return to the MainMenu or
go to DataManagement for printing, exporting, or deleting the screening record.
a. From the Screening Complete window, you may click X to return to the
MainMenu, then either power off the device or screen another infant.
b. From the Screening Complete window, you may instead click OK, which will
bring you to the DataManagement Window. You may immediately print or
export the screening record just saved. See the Chapter V: ALGO 3i Screening
Natus Medical • www.natus.com Page 10 of 12 P/N 051179E