EasyManuals Logo

Nikon XT H 225 User Manual

Nikon XT H 225
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
Important health and safety information
2 XTM0522-B1 XT H 225 X-ray System
Fusing requirements
Supply voltage/tap Recommended fuse/breaker capacity
100 V 16 A
110 V 15 A
115 V 15 A
200 V 10 A
220 V 10 A
230 V 8 A
240 V 8 A
This system is a Class A product and should not be connected to the
domestic electrical supply. In a domestic environment, this product may
cause harmonic interference. If the user wishes to connect to a
domestic supply they may be required to take adequate measures to
prevent interference.
Machines ship configured for a UK style 230 V (50/60 Hz) AC supply to
a single live conductor, a neutral conductor referenced to earth, and a
separate earth (TN-S). Other supply earth arrangements are not
supported.
The user must not configure or modify the electrical supply in any way.
Using domestic electrical fittings or inadequate wiring may lead to
fitting/wire overheating, failure, or electrical fire.
This system has high protective conductor currents
3.5 mA < I
L
< 10 mA.
It must be connected via either permanent fixed wiring or a suitable
industrial plug and socket.
1.2 Intended use
Nikon Metrology X-ray products are intended for non-destructive inspection of industrial and
electronic components using X-rays.
Do not use the system for any purpose other than its intended use.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon XT H 225 and is the answer not in the manual?

Nikon XT H 225 Specifications

General IconGeneral
BrandNikon
ModelXT H 225
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals