Maintenance
XT H 225 X-ray System XTM0522-B1 39
5.1.2 Interlock testing
The preventive maintenance schedule should include this interlock testing procedure:
Equipment Interlock testing
Name Location Activation
method
Verification
method
Test
frequency
Comment Pass/Fail
Door
Upper and
lower door
corner
Open the
door
Try to move
the
manipulator.
Should not
respond to
commands.
Weekly
Door
Upper and
lower door
corner
Open the
door
Ensure the
interlock
status is
reported to
inspect-X
properly for
the X-ray
source
interlocks as
well as the
manipulator
Weekly
Emergency
Switch OFF
(EMO)
On the
operator
table,
near
joysticks
Push the
button down
until it locks
All systems
apart from
computer are
powered
down -
pumps are
stopped,
manipulator
cannot be
moved with
controls.
Monthly