Overview
NOTE: The first-time installation of your instrument will be carried out by a PerkinElmer
Service Engineer. The following information is for reference only.
Site Requirements
For maximum stability and minimum maintenance, observe the following requirements when
choosing where to site the instrument:
A firm base free from vibration;
Enough space around and underneath the instrument for efficient air circulation, as
shown in Figure 9;
A constant temperature between 15 °C and 35 °C;
A constant relative humidity of between 20% and 80%;
An atmosphere free from dust and corrosive fumes;
Keep out of direct sunlight (illumination with diffuse lighting is ideal);
A suitable source of electrical power in the vicinity of the instrument, in this case a
proper earth-grounded 3-wire electrical outlet. The power requirements are 90250
VAC, 50/60 Hz, 250 VA.
The standard sample compartment baseplates have drain holes in them to run off spilled
liquids to the benchtop underneath the instrument. If required, place a sheet of thick filter
paper under the instrument.
Figure 9 Space requirements
74 cm (30 inches)
102 cm (41 inches)

Other manuals for PerkinElmer Lambda 1050

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 1050 and is the answer not in the manual?

PerkinElmer Lambda 1050 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 1050
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals