High-Performance Lambda Spectrometers Getting Started Guide
4. Select the options for each calibration routine.
UV/Vis Wavelength:
Select Manual set, Manual set 1 peak, Manual set 2 peaks,
Auto search 1 peak, or Auto search 2 peaks.
Manual set enables you to enter the offset and factor (without the need to
determine the values in advance).
Manual set 1 peak enables you to change the offset. A well known peak can be
shifted to an exact wavelength by entering values of Old Peak 1 and
New Peak 1.
Manual set 2 peaks enables you to change the offset and factor by specifying
the old and new values of 2 well known peaks by entering values of
Old Peak 1, New Peak 1, Old Peak 2 and New Peak 2.
Auto search 1 peak enables you to change the offset. The software performs an
automatic search for the D
2
peak at 656.1 nm. The measured peak is then shifted
to the exact wavelength.
Auto search 2 peaks enables you to change the offset and factor. The software
performs an automatic search for the D
2
peak at 656.1 nm and the peak at
0.0 nm. The measured peaks are then shifted to the exact wavelengths.
NOTE: Auto search 2 peaks is the default setting.

Other manuals for PerkinElmer Lambda 1050

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 1050 and is the answer not in the manual?

PerkinElmer Lambda 1050 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 1050
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals