11
EverFlo / EverFlo Q User Manual
Chapter 4: Alarms and Troubleshooting
Alarm and Indicators
The device has an audible alarm and three LED indicators, as shown below.
Note: All EverFlo alarms are low priority alarms. The alarm system should be veried before use and
between users by service personnel in accordance with the EverFlo service manual.
Audible Alarm / Colored LED Possible Cause Your Action
All 3 LEDs illuminate continuously
and the Audible Alarm is
sounding continuously.
The device has
detected a system
malfunction.
Immediately turn o the device,
connect to a back up oxygen
source, and call your home care
provider.
The Audible Alarm is sounding
continuously. None of the LEDs
are illuminated.
The device is turned on
but is not operating.
Often this indicates
that the device is not
plugged in or there is a
power failure.
Check the power outlet and verify
that the device is plugged in. If the
problem continues, connect to a
back up oxygen source and call
your home care provider.
Red LED illuminates continuously
and the Audible Alarm is
sounding continuously.
The device has
detected a system
malfunction.
Immediately turn o the device and
wait 5 minutes. Restart the device. If
the condition persists turn the unit
o, connect to a back up oxygen
source, and call your home care
provider.
NORMAL
Red Alarm Indicator
Yellow Check System Indicator
Green Power On Indicator