1.3 Specifications
1.3.1
Dosimeter carries out continuous measurement of the DER, DE and DE accumulation time in all
operation modes except for the computer communication mode.
1.3.2
Operating mode:
-
DER measur
ement;
- DE measurement;
- settings;
- current time indication;
- battery status indication;
- DER measurement start;
- communication with computer.
1.3.3
DER indication range
0.01
µSv/h to
12
.
0
Sv/h.
1.3.4
DER measurement range
0.1
µSv/h
to
10
.
0
Sv/h.
Limits of permissible
intrinsic
relative DER measur
e-
ment error:
(10 + К
1
/
H
2
H
) %,
where
H
− DER value in mSv/h,
K
1
− coefficient 0.0015 mSv/h,
K
2
− coefficient 0.0015 (mSv/h)
-1
.
In
the DER measurement mode dosimeter automatically calculates and indicates on LCD relative mean
square error of measurement result (statistic error) in percent at 0.95 confidence probability.
1.3.5
DE indication range:
0.001
µSv
to 24.0 Sv
1.3.6
DE measu
rement range:
-
continuous photon radiation (current):
- pulsed photon radiation (pulse duration not less than
1 ms):
0.05
µSv
to 20.0 Sv
10 µSv to 20.0 Sv
Limits of permissible
intrinsic
relative DE measur
e
ment
error:
20 %
1.3.7
Dosimeter measur
es DE accumulation time in hours and minutes and indicates it on LCD with 1
minute
resolution
1.3.8
Dosimeter allows setting two DER alarm threshold levels and two DE alarm threshold levels. Dosimeter
continuously controls the threshold levels and alerts the user with sound, light (red colour) and vibration
alarms when the threshold level is exceeded. Flash signal for the fist DER threshold and fast flash - for the
second DER threshold exceeding.
DER thresholds setting range: 0.01 µSv/h to 10.0 Sv/h
DE thresholds setting range: 1.0 µSv to 20.0 Sv
Threshold setting resolution − low order digit.
1.3.9
Photon radiation energy range
0.02 to 10.0 MeV
Energy response relative to 0.662 MeV (
137
Cs) in the
DER and DE measurement mode, no more than
-
60 % from 20
to 33 keV;
- 40 % from 33 to 48 keV;
30 % from 48 keV to 3 MeV;
50 % from 3 to 10 MeV.
1.3.10 Anisotropy of the dosimeter
for each energy does not exceed the values shown in Table 1.2 when
dosimeter is irradiated in the horizontal plane at indicated angles relative to the initial direction, and does not
exceed the values shown in Table 1.3 when the dosimeter is irradiated in the vertical plane at indicated angles
relative the initial direction.
Table 1.2
Angle of detection relative to
direction of calibration, deg
Gamma energy, MeV
Anisotropy (
), %
0
.
059
0
.
662
1
.
25
0
0
0
0
15
20
5
5
30
25
5
5
45
30
10
5
60
20
15
10
- 15
20
5
5
- 30
25
5
5
- 45
30
10
5
- 60
20
15
10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Polimaster PM1610B and is the answer not in the manual?

Polimaster PM1610B Specifications

General IconGeneral
BrandPolimaster
ModelPM1610B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals