program
DE
EN
FR
IT
ES
NL
RU
PL
SE
FI
42
Free programmable user programme
Pro-
gramme
Programme
parameter
Adjustable values
Applies in the following
incontinence type
PC1
Treatment time (min) 1min~60min
All forms of incontinence depending on the set
parameter
Work mode Work Rest/Cont.
Frequency 3 Hz~100 Hz
Rest frequency 0 Hz~4 Hz
Pulse width 50 µs~450 µs
Working hours 2 s~99 s
Ramp-up time 1 s~9 s
Ramp-down time 1 s~9 s
Rest time 2 s~99 s
Wave form Synchronous (SY) / Alternating (AL)

Table of Contents