11.1. Good Weighing Practice
To assure long-term operation and correct mass measurements, follow the
rules presented below:
Load the weighing pan steadily
avoiding mechanical shocks.
Place weighed loads centrally on the
weighing pan (eccentricity errors are
specified by EN 45501 standard, point
3.6.2.).
Do not apply concentrated force (total
load in one point).
Avoid side loading, in particular side
shocks.
11.2. Zeroing
To zero mass indication press
key. Zero indication and the following
pictograms are displayed:
and . The instrument can be zeroed only
when the indication is stable.
Indication can be zeroed only within ±2% range of maximum
capacity. If the zeroed value is greater than ±2% of the
maximum capacity, then the software displays message
<Err2>, and short sound signal is heard.