Page #47 background imageLoading...
Page #47 background image
Command Systems RIGOL
Programming Guide for DS1000B Series
2-35
PULSe Trigger
1. :TRIGger:PULSe:MODE
Command Format:
:TRIGger:PULSe:MODE <mod>
:TRIGger:PULSe:MODE?
Function:
The command sets the trigger condition. <mod> can be: +GREaterthan
(positive pulse width greater than), +LESSthan (positive pulse width less than),
+ EQUal (positive pulse width equal), -GREaterthan (negative pulse width
greater than), -LESSthan (negative pulse width less than) or –EQUal (negative
pulse width equal).
Returned Format:
The query returns +GREATER THAN or +LESS THAN, +EQUAL, -GREATER
THAN, -LESS THAN, EQUAL.
Example:
:TRIG:PULS:MODE +GRE Set the trigger condition as +GREaterthan.
:TRIG:PULS:MODE? Return +GREATER THAN.
2. :TRIGger:PULSe:WIDTh
Command Format:
:TRIGger:PULSe:WIDTh <wid>
:TRIGger:PULSe:WIDTh?
Function:
The command sets the pulse width. The <wid> range is 20ns ~10s.
Returned Format:
The query returns the value of pulse width, and the unit is s.
Example:
:TRIGr:PULS:WIDT 0.001 Set the pulse width as 1ms.
:TRIG:PULS:WIDT? Return 1.000e-003.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals