Chapter 2 Performance Verification Test RIGOL
DSG800 Performance Verification Guide 2-1
Chapter 2 Performance Verification Test
This chapter introduces the performance verification test methods and procedures of DSG800
series RF signal generator.
Note:
1) Make sure that the instrument has been warmed up for at least 40 minutes before executing
any of the following test items.
2) Reset the instrument to the factory setting before or after executing any of the following test
items.
3) The definitions of "Typical Value", "Nominal Value" and "Measured value" for this product.
Typical (typ.): the typical performance that 80 percent of the measurement results can
meet at room temperature (approximately 25℃). This data is not warranted and does not
include the measurement uncertainty.
Nominal (nom.): the expected average performance or the designed performance
attribute (for example, the 50 Ω connector). This data is not warranted and is measured
at room temperature (approximately 25℃).
Measured (meas.): the performance attribute measured during the design phase and
used to be compared with the expected performance (for example, the variation of the
amplitude drift with time). This data is not warranted and is measured at room
temperature (approximately 25℃).