RIGOL Chapter 2 Performance Verification Test
2-14 DSG800 Performance Verification Guide
FM Test
Specifications
FM
Model DSG815/DSG830
Item Condition Specification
Modulation
Accuracy
f
mod
= 1 kHz, internal modulation
< Modulation frequency deviation setting ×
2% + 20 Hz
Distortion f
mod
= 1 kHz, deviation = N
[1]
× 50 kHz < 2% (typ.)
Note: [1] The corresponding relations between the values of N and the frequency ranges are as shown in the table
below.
Frequency Range N
f < 227.5 MHz 0.25
227.5 MHz f < 455 MHz 0.125
455 MHz f < 910 MHz 0.25
910 MHz f < 1820 MHz 0.5
1820 MHz f 3000 MHz 1
Test Devices
1. Signal Analyzer × 1
2. Dual-BNC Cable × 1
3. Dual-N Cable × 1
Test Connection Diagram
Figure 2-5 FM Test Connection Diagram
Test Procedures
1. Synchronize DSG800 and the signal analyzer. Connect the RF output terminal of DSG800 with
the RF input terminal of the signal analyzer as shown in Figure 2-5.

Other manuals for Rigol DSG800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSG800 Series and is the answer not in the manual?

Rigol DSG800 Series Specifications

General IconGeneral
BrandRigol
ModelDSG800 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals