EasyManuals Logo
Home>Rigol>Portable Generator>DSG800 Series

Rigol DSG800 Series Performance Verification Guide

Rigol DSG800 Series
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
RIGOL Appendix
2 DSG800 Performance Verification Guide
Reference Value -80 dBm
DSG800 Output
Frequency
Signal Analyzer
Measurement Value A3
Global
Error
[2]
Amplitude
Accuracy
[3]
Limit
Pass/
Fail
103 kHz
1.1 dB,
0.7 (typ.)
1.03 MHz
50.03 MHz
503 MHz
1.403 GHz (Only for
DSG815)
1.903 GHz (Only for
DSG830)
2.493 GHz (Only for
DSG830)
2.903 GHz (Only for
DSG830)
Note:
[1] System Error = A2 - A1
[2] Global Error = A3 - Reference Value
[3] Amplitude Accuracy = |Global Error - System Error|
Single Sideband Phase Noise Test:
Single-sideband Phase Noise
DSG800 Output
Frequency
Offset
Measurement
Result
Limit
Pass/
Fail
100 MHz Offset 20 kHz < -100 dBc/Hz
1 GHz Offset 20 kHz < -100 dBc/Hz
3 GHz (Only for
DSG830)
Offset 20 kHz < -94 dBc/Hz