Chapter 1 Quick Start RIGOL
MSO7000/DS7000 User Guide 1-9
Function Inspection
1. Press Default on the front panel, then a prompt message "Restore default?" is
displayed. Press OK or tap OK to restore the instrument to its factory default
settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 1-10 below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-10.
Figure 1-10 To Use the Compensation Signal
4. Set the probe attenuation to 10X, and then press AUTO.
5. Observe the waveform on the display. In normal condition, the square waveform
as shown in Figure 1-11 should be displayed.
Figure 1-11 Square Waveform Signal
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation" introduced in the next section.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the
metallic part of the probe when the probe is connected to high voltage
source.
Compensation Signal Output Terminal
Ground Terminal

Table of Contents

Other manuals for Rigol MSO7024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7024 and is the answer not in the manual?

Rigol MSO7024 Specifications

General IconGeneral
BrandRigol
ModelMSO7024
CategoryTest Equipment
LanguageEnglish

Related product manuals