Chapter 4 Input/Output RIGOL
RSA3000 User Guide 4-1
Chapter 4 Input/Output
Sets the input/output interface.
Input Impedance
Sets the input impedance for voltage-to-power conversions (refer to equation (2-4)).
The default input impedance is 50 Ω. To measure a 75 Ω device, you should use a 75
Ω to 50 Ω adapter (option) supplied by RIGOL to connect the analyzer with the
system under test, and then set the input impedance to 75 Ω.
Ext Gain
Compensates for gain or loss in the measurement system outside the instrument.
Remarks:
The value does not affect the trace position, but will modify the parameters such
as reference level readout and the marker amplitude readout.
You can use the numeric keys, the knob, and the arrow keys on the front panel
to modify this parameter; also you can modify it on the touchscreen. For details,
refer to descriptions in "Parameter Setting".
Table 4-1 External Gain
Parameter
Remarks
Default
0 dB
Range
-120 dB to 120 dB
Unit
dB
Knob Step
1 dB
Left/Right Arrow Key Step
Up/Down Arrow Key Step
5 dB
Ext Trigger2
Sets the external trigger 2 [TRIGGER IN/OUT] to be the "input" or "output"
interface.

Table of Contents

Other manuals for Rigol RSA3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3030 and is the answer not in the manual?

Rigol RSA3030 Specifications

General IconGeneral
BrandRigol
ModelRSA3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals