Chapter 2 Performance Verification Test RIGOL
RSA3000 Performance Verification Guide
2-11
Test Record Form
PA Off:
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/F
ail
100 kHz
100 kHz
to 3.0
GHz: <0.7
dB, <0.5
dB (typical)
3.0 GHz
to 4.5
GHz:<0.9
dB, <0.5
dB (typical)
500 MHz
2 GHz
4 GHz
4.499995
GHz
PA On:
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/F
ail
100 kHz
100 kHz
to 3.0
GHz: <1.0
dB, <0.5
dB (typical)
3.0 GHz
to 4.5
GHz:<1.2
dB, <0.5
500 MHz
2 GHz
4 GHz
4.499995
GHz

Other manuals for Rigol RSA5065N-OCXO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA5065N-OCXO and is the answer not in the manual?

Rigol RSA5065N-OCXO Specifications

General IconGeneral
BrandRigol
ModelRSA5065N-OCXO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals